Family Laser · Burns
Product: Laser therapy device · Pannonfitt · CDU23, JG044961
Category: Electrical appliances and equipment · Hungary
Hazard: The battery management system lacks a cell temperature monitoring device. The product can overheat and cause burns. The product does not comply with the requirements of the Low Voltage Directive nor with the European standards EN 62133 and EN 60335.
Official action: Type of economic operator to whom the measure(s) were ordered: Importer Category of measure(s): Withdrawal of the product from the market Date of entry into force: 23/06/2025Type of economic operator to whom the measure(s) were ordered: Importer Category of measure(s): Recall of the product from end users Date of entry into force: 23/06/2025