Personal Face Mask N95 · Health risk / other
Product: Particle filter mask · Panodyne · FBN95M-U
Category: Protective equipment · Malta
Hazard: The product bears a CE marking but its filtering capacity has not been tested by a relevant competent European conformity assessment body.Additionally, the particle/filter retention of the material is insufficient (measured values: as low as 39%), the total filtration capacity of the mask is insufficient (measured mean values: as low as 75%).Consequently, the product does not fulfil the health and
Official action: Type of economic operator to whom the measure(s) were ordered: Other Category of measure(s): Recall of the product from end users Date of entry into force: 04/05/2021